Test Date: 2016-04-11 17:49
Analysis date: 2016-05-26 13:55
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 54.65 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.38
Rate Rate 287.60 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 109.64 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.11
Rate Rate 436.80 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 137.52 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.89
Rate Rate 232.30 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 156.58 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.03
Rate Rate 148.40 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 46.281 ± 0.151 nElectrons/VCal
Offset Offset -217.012 ± 19.072 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 287.60 kHz/cm²
NTrigs N Trig 5916031 Trigger
NHits N Hits 26545 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 436.80 kHz/cm²
NTrigs N Trig 5916365 Trigger
NHits N Hits 40315 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 232.30 kHz/cm²
NTrigs N Trig 5916113 Trigger
NHits N Hits 21443 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 148.40 kHz/cm²
NTrigs N Trig 5916144 Trigger
NHits N Hits 13692 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master