Test Date: 2015-07-07 08:07
Chip 12
X-ray Calibration - Target Zn Method Spectrum Chip C12
Center Center of Peak 47.65 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 30.66
Rate Rate 283.70 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C12
Center Center of Peak 96.42 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 12.46
Rate Rate 451.60 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C12
Center Center of Peak 123.21 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 7.68
Rate Rate 243.30 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C12
Center Center of Peak 141.5 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 8.09
Rate Rate 152.00 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C12
Slope Slope 48.194 ± 0.156 nElectrons/VCal
Offset Offset 209.917 ± 17.186 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C12
Rate Rate 283.70 kHz/cm²
NTrigs N Trig 5916644 Trigger
NHits N Hits 26183 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C12
Rate Rate 451.60 kHz/cm²
NTrigs N Trig 5916569 Trigger
NHits N Hits 41681 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C12
Rate Rate 243.30 kHz/cm²
NTrigs N Trig 5916456 Trigger
NHits N Hits 22455 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C12
Rate Rate 152.00 kHz/cm²
NTrigs N Trig 5916385 Trigger
NHits N Hits 14030 Hits