Test Date: 2015-07-07 08:07
Chip 10
X-ray Calibration - Target Zn Method Spectrum Chip C10
Center Center of Peak 47.47 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 15.69
Rate Rate 336.70 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C10
Center Center of Peak 100.6 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 6.56
Rate Rate 479.90 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C10
Center Center of Peak 129.09 ± 0.12 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 5.6
Rate Rate 249.50 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C10
Center Center of Peak 147.93 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.11
Rate Rate 156.00 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C10
Slope Slope 45.712 ± 0.149 nElectrons/VCal
Offset Offset 256.517 ± 17.452 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C10
Rate Rate 336.70 kHz/cm²
NTrigs N Trig 5916644 Trigger
NHits N Hits 31079 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C10
Rate Rate 479.90 kHz/cm²
NTrigs N Trig 5916569 Trigger
NHits N Hits 44297 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C10
Rate Rate 249.50 kHz/cm²
NTrigs N Trig 5916456 Trigger
NHits N Hits 23031 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C10
Rate Rate 156.00 kHz/cm²
NTrigs N Trig 5916385 Trigger
NHits N Hits 14398 Hits