Test Date: 2015-07-07 08:07
Chip 1
X-ray Calibration - Target Zn Method Spectrum Chip C1
Center Center of Peak 49.48 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 29.05
Rate Rate 237.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C1
Center Center of Peak 100.82 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 13.92
Rate Rate 378.70 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C1
Center Center of Peak 127.93 ± 0.13 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 11.36
Rate Rate 220.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C1
Center Center of Peak 146.83 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 9.22
Rate Rate 142.80 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C1
Slope Slope 47.866 ± 0.087 nElectrons/VCal
Offset Offset 30.751 ± 7.99 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C1
Rate Rate 237.20 kHz/cm²
NTrigs N Trig 5916644 Trigger
NHits N Hits 21896 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C1
Rate Rate 378.70 kHz/cm²
NTrigs N Trig 5916569 Trigger
NHits N Hits 34955 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C1
Rate Rate 220.80 kHz/cm²
NTrigs N Trig 5916456 Trigger
NHits N Hits 20379 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C1
Rate Rate 142.80 kHz/cm²
NTrigs N Trig 5916385 Trigger
NHits N Hits 13178 Hits