Test Date: 2015-07-07 08:07
Chip 0
X-ray Calibration - Target Zn Method Spectrum Chip C0
Center Center of Peak 43.73 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 16.84
Rate Rate 252.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C0
Center Center of Peak 92.02 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 8.3
Rate Rate 389.60 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C0
Center Center of Peak 117.39 ± 0.12 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 5.77
Rate Rate 222.40 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C0
Center Center of Peak 134.57 ± 0.16 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 5.73
Rate Rate 143.30 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C0
Slope Slope 50.922 ± 0.087 nElectrons/VCal
Offset Offset 172.209 ± 8.132 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C0
Rate Rate 252.10 kHz/cm²
NTrigs N Trig 5916644 Trigger
NHits N Hits 23268 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C0
Rate Rate 389.60 kHz/cm²
NTrigs N Trig 5916569 Trigger
NHits N Hits 35964 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C0
Rate Rate 222.40 kHz/cm²
NTrigs N Trig 5916456 Trigger
NHits N Hits 20528 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C0
Rate Rate 143.30 kHz/cm²
NTrigs N Trig 5916385 Trigger
NHits N Hits 13228 Hits