Test Date: 2015-04-08 11:04
Chip 5
X-ray Calibration - Target Cu Method Spectrum Chip C5
Center Center of Peak 43.89 ± 0.32 Vcal
TargetEnergy Energy of target Cu 8047.82 eV
TargetNElectrons Energy of target Cu 2235.51 nElectrons
Chi2PerNDF Chi^2 per NDF 44.29
Rate Rate 341.80 kHz/cm²
X-ray Calibration - Target Zn Method Spectrum Chip C5
Center Center of Peak 49.85 ± 0.16 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 49.03
Rate Rate 385.20 kHz/cm²
X-ray Calibration - Target Br Method Spectrum Chip C5
Center Center of Peak 65.14 ± 0.08 Vcal
TargetEnergy Energy of target Br 11924.36 eV
TargetNElectrons Energy of target Br 3312.32 nElectrons
Chi2PerNDF Chi^2 per NDF 34.94
Rate Rate 590.90 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C5
Center Center of Peak 94.7 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 19.84
Rate Rate 570.20 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C5
Center Center of Peak 120.49 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 11.68
Rate Rate 294.00 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C5
Center Center of Peak 137.91 ± 0.16 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.35
Rate Rate 187.30 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C5
Slope Slope 50.205 ± 0.192 nElectrons/VCal
Offset Offset 102.274 ± 21.398 nElectrons
X-ray Hit Map - Target Cu Method Spectrum Chip C5
Rate Rate 341.80 kHz/cm²
NTrigs N Trig 2956313 Trigger
NHits N Hits 15762 Hits
X-ray Hit Map - Target Zn Method Spectrum Chip C5
Rate Rate 385.20 kHz/cm²
NTrigs N Trig 2956201 Trigger
NHits N Hits 17762 Hits
X-ray Hit Map - Target Br Method Spectrum Chip C5
Rate Rate 590.90 kHz/cm²
NTrigs N Trig 2956067 Trigger
NHits N Hits 27247 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C5
Rate Rate 570.20 kHz/cm²
NTrigs N Trig 2956113 Trigger
NHits N Hits 26294 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C5
Rate Rate 294.00 kHz/cm²
NTrigs N Trig 3750020 Trigger
NHits N Hits 17197 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C5
Rate Rate 187.30 kHz/cm²
NTrigs N Trig 2956395 Trigger
NHits N Hits 8636 Hits