Test Date: 2016-10-10 09:55
Analysis date: 2016-10-10 11:33
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center |
Center of Peak |
53.33 ± 0.07 |
Vcal |
TargetEnergy |
Energy of target Zn |
8638.91 |
eV |
TargetNElectrons |
Energy of target Zn |
2399.7 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
1.6 |
|
Rate |
Rate |
310.00 |
kHz/cm² |
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center |
Center of Peak |
104.25 ± 0.07 |
Vcal |
TargetEnergy |
Energy of target Mo |
17479.37 |
eV |
TargetNElectrons |
Energy of target Mo |
4855.38 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
2.3 |
|
Rate |
Rate |
450.60 |
kHz/cm² |
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center |
Center of Peak |
132.37 ± 0.1 |
Vcal |
TargetEnergy |
Energy of target Ag |
22162.92 |
eV |
TargetNElectrons |
Energy of target Ag |
6156.37 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
3.76 |
|
Rate |
Rate |
242.20 |
kHz/cm² |
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center |
Center of Peak |
150.36 ± 0.14 |
Vcal |
TargetEnergy |
Energy of target Sn |
25271.36 |
eV |
TargetNElectrons |
Energy of target Sn |
7019.82 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
3.98 |
|
Rate |
Rate |
149.60 |
kHz/cm² |
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope |
Slope |
46.715 ± 0.14 |
nElectrons/VCal |
Offset |
Offset |
-16.76 ± 16.791 |
nElectrons |
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate |
Rate |
310.00 |
kHz/cm² |
NTrigs |
N Trig |
5913974 |
Trigger |
NHits |
N Hits |
28601 |
Hits |
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate |
Rate |
450.60 |
kHz/cm² |
NTrigs |
N Trig |
5913773 |
Trigger |
NHits |
N Hits |
41573 |
Hits |
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate |
Rate |
242.20 |
kHz/cm² |
NTrigs |
N Trig |
5914000 |
Trigger |
NHits |
N Hits |
22343 |
Hits |
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate |
Rate |
149.60 |
kHz/cm² |
NTrigs |
N Trig |
5914196 |
Trigger |
NHits |
N Hits |
13802 |
Hits |
MoReWeb-v1.0.4-4-g297680c on branch psi46master