Test Date: 2016-08-03 11:22
Analysis date: 2016-08-04 09:48
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 53.73 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.33
Rate Rate 240.30 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 107.3 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.59
Rate Rate 401.80 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 135.72 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.53
Rate Rate 229.60 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 153.89 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.01
Rate Rate 148.90 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 45.825 ± 0.07 nElectrons/VCal
Offset Offset -62.28 ± 6.881 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 240.30 kHz/cm²
NTrigs N Trig 5918134 Trigger
NHits N Hits 22182 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 401.80 kHz/cm²
NTrigs N Trig 5918707 Trigger
NHits N Hits 37097 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 229.60 kHz/cm²
NTrigs N Trig 5918327 Trigger
NHits N Hits 21197 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 148.90 kHz/cm²
NTrigs N Trig 5919265 Trigger
NHits N Hits 13750 Hits
MoReWeb-v1.0.2-10-gbc52a17 on branch psi46master