Test Date: 2016-01-19 10:15
Analysis date: 2016-05-26 08:14
Chip 13
X-ray Calibration - Target Zn Method Spectrum Chip C13
Center Center of Peak 50.72 ± 0.12 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 60.93
Rate Rate 291.00 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C13
Center Center of Peak 100.58 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 67.58
Rate Rate 442.90 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C13
Center Center of Peak 127.12 ± 0.13 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 79.0
Rate Rate 332.00 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C13
Center Center of Peak 144.22 ± 0.16 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 70.14
Rate Rate 256.50 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C13
Slope Slope 49.19 ± 0.11 nElectrons/VCal
Offset Offset -94.032 ± 10.691 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C13
Rate Rate 291.00 kHz/cm²
NTrigs N Trig 5914074 Trigger
NHits N Hits 26847 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C13
Rate Rate 442.90 kHz/cm²
NTrigs N Trig 5913885 Trigger
NHits N Hits 40860 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C13
Rate Rate 332.00 kHz/cm²
NTrigs N Trig 5913947 Trigger
NHits N Hits 30625 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C13
Rate Rate 256.50 kHz/cm²
NTrigs N Trig 5914004 Trigger
NHits N Hits 23665 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master