Test Date: 2015-08-11 12:52
Analysis date: 2016-05-26 00:47
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 51.98 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.45
Rate Rate 219.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 104.32 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.42
Rate Rate 365.30 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 132.23 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.16
Rate Rate 215.00 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 150.19 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.43
Rate Rate 142.00 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 46.842 ± 0.066 nElectrons/VCal
Offset Offset -34.299 ± 6.165 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 219.20 kHz/cm²
NTrigs N Trig 5920914 Trigger
NHits N Hits 20243 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 365.30 kHz/cm²
NTrigs N Trig 5920059 Trigger
NHits N Hits 33732 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 215.00 kHz/cm²
NTrigs N Trig 5920013 Trigger
NHits N Hits 19858 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 142.00 kHz/cm²
NTrigs N Trig 5920526 Trigger
NHits N Hits 13111 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master