Test Date: 2015-07-07 08:06
Analysis date: 2016-05-25 21:25
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 47.67 ± 0.13 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 23.3
Rate Rate 225.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 104.23 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 11.9
Rate Rate 376.40 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 134.23 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 8.81
Rate Rate 214.90 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 154.11 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.4
Rate Rate 139.70 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 43.387 ± 0.124 nElectrons/VCal
Offset Offset 333.088 ± 15.376 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 225.10 kHz/cm²
NTrigs N Trig 5916700 Trigger
NHits N Hits 20777 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 376.40 kHz/cm²
NTrigs N Trig 5916774 Trigger
NHits N Hits 34739 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 214.90 kHz/cm²
NTrigs N Trig 5913533 Trigger
NHits N Hits 19825 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 139.70 kHz/cm²
NTrigs N Trig 5917045 Trigger
NHits N Hits 12896 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master