Test Date: 2019-05-03 10:27
Analysis date: 2019-05-03 15:24
Chip 6
X-ray Calibration - Target Zn Method Spectrum Chip C6
Center Center of Peak 63.85 ± 0.16 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.2
Rate Rate 182.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C6
Center Center of Peak 122.38 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.36
Rate Rate 416.20 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C6
Center Center of Peak 151.89 ± 0.14 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.14
Rate Rate 240.10 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C6
Center Center of Peak 171.81 ± 0.19 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 2.48
Rate Rate 151.90 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C6
Slope Slope 43.885 ± 0.164 nElectrons/VCal
Offset Offset -514.447 ± 22.669 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C6
Rate Rate 182.20 kHz/cm²
NTrigs N Trig 5915949 Trigger
NHits N Hits 16819 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C6
Rate Rate 416.20 kHz/cm²
NTrigs N Trig 5915789 Trigger
NHits N Hits 38409 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C6
Rate Rate 240.10 kHz/cm²
NTrigs N Trig 5915976 Trigger
NHits N Hits 22162 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C6
Rate Rate 151.90 kHz/cm²
NTrigs N Trig 5915857 Trigger
NHits N Hits 14018 Hits
MoReWeb-v1.1.0 on branch AnalyzeNewXPixelAlive