Test Date: 2015-11-04 10:51
Analysis date: 2015-11-23 15:53
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 52.79 ± 0.11 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.38
Rate Rate 319.40 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 104.04 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.66
Rate Rate 468.90 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 132.11 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 2.86
Rate Rate 244.70 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 150.14 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.46
Rate Rate 155.80 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 46.776 ± 0.162 nElectrons/VCal
Offset Offset -13.736 ± 19.891 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 319.40 kHz/cm²
NTrigs N Trig 5920158 Trigger
NHits N Hits 29498 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 468.90 kHz/cm²
NTrigs N Trig 5920582 Trigger
NHits N Hits 43304 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 244.70 kHz/cm²
NTrigs N Trig 5920220 Trigger
NHits N Hits 22597 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 155.80 kHz/cm²
NTrigs N Trig 5921139 Trigger
NHits N Hits 14387 Hits
MoReWeb-v0.6.7-72-g43ac54d on branch dev-v0.7.0