Test Date: 2016-08-18 11:13
Analysis date: 2016-08-18 13:31
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 51.99 ± 0.13 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.09
Rate Rate 118.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 103.23 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.43
Rate Rate 363.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 130.65 ± 0.12 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.16
Rate Rate 213.70 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 149.25 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 2.89
Rate Rate 139.70 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 47.782 ± 0.106 nElectrons/VCal
Offset Offset -81.418 ± 10.909 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 118.10 kHz/cm²
NTrigs N Trig 5921889 Trigger
NHits N Hits 10907 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 363.50 kHz/cm²
NTrigs N Trig 5922929 Trigger
NHits N Hits 33590 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 213.70 kHz/cm²
NTrigs N Trig 5921809 Trigger
NHits N Hits 19742 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 139.70 kHz/cm²
NTrigs N Trig 5921238 Trigger
NHits N Hits 12900 Hits
MoReWeb-v1.0.2-10-gbc52a17 on branch psi46master