Test Date: 2015-12-11 15:01
Analysis date: 2015-12-15 10:00
Chip 2
X-ray Calibration - Target Zn Method Spectrum Chip C2
Center Center of Peak 56.64 ± 0.2 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 0.83
Rate Rate 76.80 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C2
Center Center of Peak 113.54 ± 0.11 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.52
Rate Rate 259.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C2
Center Center of Peak 143.92 ± 0.14 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 2.87
Rate Rate 155.70 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C2
Center Center of Peak 163.58 ± 0.19 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.39
Rate Rate 103.20 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C2
Slope Slope 43.038 ± 0.119 nElectrons/VCal
Offset Offset -34.374 ± 14.07 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C2
Rate Rate 76.80 kHz/cm²
NTrigs N Trig 5921517 Trigger
NHits N Hits 7096 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C2
Rate Rate 259.50 kHz/cm²
NTrigs N Trig 5922269 Trigger
NHits N Hits 23976 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C2
Rate Rate 155.70 kHz/cm²
NTrigs N Trig 5922329 Trigger
NHits N Hits 14384 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C2
Rate Rate 103.20 kHz/cm²
NTrigs N Trig 5922180 Trigger
NHits N Hits 9534 Hits
MoReWeb-v0.6.7-111-gcc5e703 on branch 20151208_Readback