Test Date: 2015-10-26 18:02
Analysis date: 2015-10-26 19:56
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 47.24 ± 0.15 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.09
Rate Rate 112.60 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 93.46 ± 0.14 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.05
Rate Rate 334.80 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 117.35 ± 0.18 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 2.63
Rate Rate 204.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 133.76 ± 0.23 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 2.25
Rate Rate 132.80 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 53.904 ± 0.335 nElectrons/VCal
Offset Offset -179.903 ± 36.656 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 112.60 kHz/cm²
NTrigs N Trig 5924901 Trigger
NHits N Hits 10404 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 334.80 kHz/cm²
NTrigs N Trig 5924167 Trigger
NHits N Hits 30937 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 204.80 kHz/cm²
NTrigs N Trig 5925119 Trigger
NHits N Hits 18934 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 132.80 kHz/cm²
NTrigs N Trig 5925477 Trigger
NHits N Hits 12278 Hits
MoReWeb-v0.6.7-11-ga66536c on branch psi46master