Test Date: 2015-07-07 08:07
Chip 9
X-ray Calibration - Target Zn Method Spectrum Chip C9
Center Center of Peak 50.99 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 22.0
Rate Rate 293.00 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C9
Center Center of Peak 104.15 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 8.95
Rate Rate 449.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C9
Center Center of Peak 132.25 ± 0.13 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 8.63
Rate Rate 240.30 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C9
Center Center of Peak 152.11 ± 0.16 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.84
Rate Rate 152.70 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C9
Slope Slope 46.217 ± 0.081 nElectrons/VCal
Offset Offset 42.731 ± 7.695 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C9
Rate Rate 293.00 kHz/cm²
NTrigs N Trig 5916644 Trigger
NHits N Hits 27048 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C9
Rate Rate 449.10 kHz/cm²
NTrigs N Trig 5916569 Trigger
NHits N Hits 41454 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C9
Rate Rate 240.30 kHz/cm²
NTrigs N Trig 5916456 Trigger
NHits N Hits 22178 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C9
Rate Rate 152.70 kHz/cm²
NTrigs N Trig 5916385 Trigger
NHits N Hits 14092 Hits