Test Date: 2015-07-07 08:07
Chip 6
X-ray Calibration - Target Zn Method Spectrum Chip C6
Center Center of Peak 51.43 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 12.65
Rate Rate 294.80 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C6
Center Center of Peak 103.37 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 6.93
Rate Rate 444.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C6
Center Center of Peak 130.38 ± 0.12 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 7.65
Rate Rate 241.40 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C6
Center Center of Peak 148.8 ± 0.16 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.23
Rate Rate 152.90 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C6
Slope Slope 47.806 ± 0.168 nElectrons/VCal
Offset Offset -84.84 ± 19.909 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C6
Rate Rate 294.80 kHz/cm²
NTrigs N Trig 5916644 Trigger
NHits N Hits 27211 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C6
Rate Rate 444.10 kHz/cm²
NTrigs N Trig 5916569 Trigger
NHits N Hits 40990 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C6
Rate Rate 241.40 kHz/cm²
NTrigs N Trig 5916456 Trigger
NHits N Hits 22284 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C6
Rate Rate 152.90 kHz/cm²
NTrigs N Trig 5916385 Trigger
NHits N Hits 14113 Hits