Test Date: 2015-07-07 08:07
Chip 9
X-ray Calibration - Target Zn Method Spectrum Chip C9
Center Center of Peak 52.16 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 14.32
Rate Rate 304.60 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C9
Center Center of Peak 103.59 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 8.02
Rate Rate 435.20 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C9
X-ray Calibration - Target Sn Method Spectrum Chip C9
Center Center of Peak 149.23 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 5.66
Rate Rate 151.50 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C9
X-ray Hit Map - Target Zn Method Spectrum Chip C9
Rate Rate 304.60 kHz/cm²
NTrigs N Trig 5920420 Trigger
NHits N Hits 28134 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C9
Rate Rate 435.20 kHz/cm²
NTrigs N Trig 5920389 Trigger
NHits N Hits 40195 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C9
X-ray Hit Map - Target Sn Method Spectrum Chip C9
Rate Rate 151.50 kHz/cm²
NTrigs N Trig 5919695 Trigger
NHits N Hits 13992 Hits