Test Date: 2015-07-07 08:07
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center |
Center of Peak |
50.53 ± 0.07 |
Vcal |
TargetEnergy |
Energy of target Zn |
8638.91 |
eV |
TargetNElectrons |
Energy of target Zn |
2399.7 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
18.64 |
|
Rate |
Rate |
296.00 |
kHz/cm² |
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center |
Center of Peak |
98.05 ± 0.08 |
Vcal |
TargetEnergy |
Energy of target Mo |
17479.37 |
eV |
TargetNElectrons |
Energy of target Mo |
4855.38 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
8.45 |
|
Rate |
Rate |
434.90 |
kHz/cm² |
X-ray Calibration - Target Ag Method Spectrum Chip C8
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center |
Center of Peak |
141.06 ± 0.14 |
Vcal |
TargetEnergy |
Energy of target Sn |
25271.36 |
eV |
TargetNElectrons |
Energy of target Sn |
7019.82 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
4.61 |
|
Rate |
Rate |
160.10 |
kHz/cm² |
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate |
Rate |
296.00 |
kHz/cm² |
NTrigs |
N Trig |
5920420 |
Trigger |
NHits |
N Hits |
27339 |
Hits |
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate |
Rate |
434.90 |
kHz/cm² |
NTrigs |
N Trig |
5920389 |
Trigger |
NHits |
N Hits |
40166 |
Hits |
X-ray Hit Map - Target Ag Method Spectrum Chip C8
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate |
Rate |
160.10 |
kHz/cm² |
NTrigs |
N Trig |
5919695 |
Trigger |
NHits |
N Hits |
14782 |
Hits |