Test Date: 2015-07-07 08:07
Chip 3
X-ray Calibration - Target Zn Method Spectrum Chip C3
Center Center of Peak 48.53 ± 0.15 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 72.8
Rate Rate 300.50 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C3
Center Center of Peak 94.51 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 22.99
Rate Rate 431.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C3
X-ray Calibration - Target Sn Method Spectrum Chip C3
Center Center of Peak 138.54 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 9.59
Rate Rate 154.40 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C3
X-ray Hit Map - Target Zn Method Spectrum Chip C3
Rate Rate 300.50 kHz/cm²
NTrigs N Trig 5920420 Trigger
NHits N Hits 27753 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C3
Rate Rate 431.10 kHz/cm²
NTrigs N Trig 5920389 Trigger
NHits N Hits 39819 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C3
X-ray Hit Map - Target Sn Method Spectrum Chip C3
Rate Rate 154.40 kHz/cm²
NTrigs N Trig 5919695 Trigger
NHits N Hits 14261 Hits