Test Date: 2015-07-07 08:07
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center |
Center of Peak |
45.15 ± 0.08 |
Vcal |
TargetEnergy |
Energy of target Zn |
8638.91 |
eV |
TargetNElectrons |
Energy of target Zn |
2399.7 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
18.92 |
|
Rate |
Rate |
248.60 |
kHz/cm² |
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center |
Center of Peak |
93.55 ± 0.09 |
Vcal |
TargetEnergy |
Energy of target Mo |
17479.37 |
eV |
TargetNElectrons |
Energy of target Mo |
4855.38 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
9.23 |
|
Rate |
Rate |
367.80 |
kHz/cm² |
X-ray Calibration - Target Ag Method Spectrum Chip C14
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center |
Center of Peak |
136.38 ± 0.13 |
Vcal |
TargetEnergy |
Energy of target Sn |
25271.36 |
eV |
TargetNElectrons |
Energy of target Sn |
7019.82 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
5.17 |
|
Rate |
Rate |
146.40 |
kHz/cm² |
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate |
Rate |
248.60 |
kHz/cm² |
NTrigs |
N Trig |
5920420 |
Trigger |
NHits |
N Hits |
22956 |
Hits |
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate |
Rate |
367.80 |
kHz/cm² |
NTrigs |
N Trig |
5920389 |
Trigger |
NHits |
N Hits |
33965 |
Hits |
X-ray Hit Map - Target Ag Method Spectrum Chip C14
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate |
Rate |
146.40 |
kHz/cm² |
NTrigs |
N Trig |
5919695 |
Trigger |
NHits |
N Hits |
13517 |
Hits |