Test Date: 2015-07-07 08:07
Chip 13
X-ray Calibration - Target Zn Method Spectrum Chip C13
Center Center of Peak 52.27 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 8.94
Rate Rate 191.40 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C13
Center Center of Peak 103.84 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 6.05
Rate Rate 313.90 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C13
X-ray Calibration - Target Sn Method Spectrum Chip C13
Center Center of Peak 149.66 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.2
Rate Rate 122.70 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C13
X-ray Hit Map - Target Zn Method Spectrum Chip C13
Rate Rate 191.40 kHz/cm²
NTrigs N Trig 5920420 Trigger
NHits N Hits 17679 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C13
Rate Rate 313.90 kHz/cm²
NTrigs N Trig 5920389 Trigger
NHits N Hits 28988 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C13
X-ray Hit Map - Target Sn Method Spectrum Chip C13
Rate Rate 122.70 kHz/cm²
NTrigs N Trig 5919695 Trigger
NHits N Hits 11330 Hits