Test Date: 2015-07-07 08:07
Chip 11
X-ray Calibration - Target Zn Method Spectrum Chip C11
Center Center of Peak 51.85 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 26.06
Rate Rate 304.30 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C11
Center Center of Peak 99.11 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 9.72
Rate Rate 443.30 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C11
X-ray Calibration - Target Sn Method Spectrum Chip C11
Center Center of Peak 143.51 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.33
Rate Rate 155.80 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C11
X-ray Hit Map - Target Zn Method Spectrum Chip C11
Rate Rate 304.30 kHz/cm²
NTrigs N Trig 5920420 Trigger
NHits N Hits 28105 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C11
Rate Rate 443.30 kHz/cm²
NTrigs N Trig 5920389 Trigger
NHits N Hits 40942 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C11
X-ray Hit Map - Target Sn Method Spectrum Chip C11
Rate Rate 155.80 kHz/cm²
NTrigs N Trig 5919695 Trigger
NHits N Hits 14384 Hits