Test Date: 2015-05-20 17:05
Chip 1
X-ray Calibration - Target Zn Method Spectrum Chip C1
Center |
Center of Peak |
55.75 ± 0.09 |
Vcal |
TargetEnergy |
Energy of target Zn |
8638.91 |
eV |
TargetNElectrons |
Energy of target Zn |
2399.7 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
5.28 |
|
Rate |
Rate |
266.70 |
kHz/cm² |
X-ray Calibration - Target Br Method Spectrum Chip C1
Center |
Center of Peak |
79.42 ± 0.1 |
Vcal |
TargetEnergy |
Energy of target Br |
11924.36 |
eV |
TargetNElectrons |
Energy of target Br |
3312.32 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
5.59 |
|
Rate |
Rate |
366.50 |
kHz/cm² |
X-ray Calibration - Target Mo Method Spectrum Chip C1
Center |
Center of Peak |
116.63 ± 0.09 |
Vcal |
TargetEnergy |
Energy of target Mo |
17479.37 |
eV |
TargetNElectrons |
Energy of target Mo |
4855.38 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
4.8 |
|
Rate |
Rate |
410.40 |
kHz/cm² |
X-ray Calibration - Target Ag Method Spectrum Chip C1
Center |
Center of Peak |
148.26 ± 0.11 |
Vcal |
TargetEnergy |
Energy of target Ag |
22162.92 |
eV |
TargetNElectrons |
Energy of target Ag |
6156.37 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
5.74 |
|
Rate |
Rate |
237.60 |
kHz/cm² |
X-ray Calibration - Target Sn Method Spectrum Chip C1
Center |
Center of Peak |
168.04 ± 0.16 |
Vcal |
TargetEnergy |
Energy of target Sn |
25271.36 |
eV |
TargetNElectrons |
Energy of target Sn |
7019.82 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
7.2 |
|
Rate |
Rate |
152.50 |
kHz/cm² |
X-ray Calibration - Vcal Calibration Method Spectrum Chip C1
Slope |
Slope |
40.544 ± 0.059 |
nElectrons/VCal |
Offset |
Offset |
136.104 ± 6.432 |
nElectrons |
X-ray Hit Map - Target Zn Method Spectrum Chip C1
Rate |
Rate |
266.70 |
kHz/cm² |
NTrigs |
N Trig |
5910292 |
Trigger |
NHits |
N Hits |
24586 |
Hits |
X-ray Hit Map - Target Br Method Spectrum Chip C1
Rate |
Rate |
366.50 |
kHz/cm² |
NTrigs |
N Trig |
5909321 |
Trigger |
NHits |
N Hits |
33790 |
Hits |
X-ray Hit Map - Target Mo Method Spectrum Chip C1
Rate |
Rate |
410.40 |
kHz/cm² |
NTrigs |
N Trig |
5909179 |
Trigger |
NHits |
N Hits |
37829 |
Hits |
X-ray Hit Map - Target Ag Method Spectrum Chip C1
Rate |
Rate |
237.60 |
kHz/cm² |
NTrigs |
N Trig |
6065661 |
Trigger |
NHits |
N Hits |
22480 |
Hits |
X-ray Hit Map - Target Sn Method Spectrum Chip C1
Rate |
Rate |
152.50 |
kHz/cm² |
NTrigs |
N Trig |
5335050 |
Trigger |
NHits |
N Hits |
12693 |
Hits |