Test Date: 2015-06-05 18:06
Chip 2
X-ray Calibration - Target Zn Method Spectrum Chip C2
Center Center of Peak 52.8 ± 0.14 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 31.1
Rate Rate 180.60 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C2
Center Center of Peak 102.69 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 24.83
Rate Rate 306.70 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C2
Center Center of Peak 130.46 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 13.11
Rate Rate 174.70 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C2
Center Center of Peak 148.69 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 9.32
Rate Rate 114.10 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C2
Slope Slope 46.997 ± 0.151 nElectrons/VCal
Offset Offset 28.566 ± 18.086 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C2
Rate Rate 180.60 kHz/cm²
NTrigs N Trig 5913115 Trigger
NHits N Hits 16659 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C2
Rate Rate 306.70 kHz/cm²
NTrigs N Trig 5913243 Trigger
NHits N Hits 28296 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C2
Rate Rate 174.70 kHz/cm²
NTrigs N Trig 5913128 Trigger
NHits N Hits 16119 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C2
Rate Rate 114.10 kHz/cm²
NTrigs N Trig 5912980 Trigger
NHits N Hits 10521 Hits