Test Date: 2015-06-05 18:06
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 57.75 ± 0.11 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 40.4
Rate Rate 250.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 109.72 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 26.96
Rate Rate 391.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 138.23 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 16.62
Rate Rate 225.20 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 157.39 ± 0.12 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 12.43
Rate Rate 145.00 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 45.462 ± 0.123 nElectrons/VCal
Offset Offset -131.906 ± 15.655 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 250.10 kHz/cm²
NTrigs N Trig 5913115 Trigger
NHits N Hits 23072 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 391.50 kHz/cm²
NTrigs N Trig 5913243 Trigger
NHits N Hits 36114 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 225.20 kHz/cm²
NTrigs N Trig 5913128 Trigger
NHits N Hits 20774 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 145.00 kHz/cm²
NTrigs N Trig 5912980 Trigger
NHits N Hits 13373 Hits