Test Date: 2015-06-05 18:06
Chip 1
X-ray Calibration - Target Zn Method Spectrum Chip C1
Center Center of Peak 54.08 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 39.64
Rate Rate 251.50 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C1
Center Center of Peak 104.37 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 24.89
Rate Rate 387.40 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C1
Center Center of Peak 132.12 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 15.97
Rate Rate 226.20 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C1
Center Center of Peak 150.25 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 10.33
Rate Rate 144.60 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C1
Slope Slope 47.09 ± 0.137 nElectrons/VCal
Offset Offset -60.344 ± 16.511 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C1
Rate Rate 251.50 kHz/cm²
NTrigs N Trig 5913115 Trigger
NHits N Hits 23198 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C1
Rate Rate 387.40 kHz/cm²
NTrigs N Trig 5913243 Trigger
NHits N Hits 35736 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C1
Rate Rate 226.20 kHz/cm²
NTrigs N Trig 5913128 Trigger
NHits N Hits 20870 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C1
Rate Rate 144.60 kHz/cm²
NTrigs N Trig 5912980 Trigger
NHits N Hits 13342 Hits