Test Date: 2015-06-05 18:06
Chip 0
X-ray Calibration - Target Zn Method Spectrum Chip C0
Center Center of Peak 53.74 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 43.74
Rate Rate 246.30 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C0
Center Center of Peak 104.02 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 29.73
Rate Rate 387.70 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C0
Center Center of Peak 131.01 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 15.43
Rate Rate 223.60 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C0
Center Center of Peak 149.21 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 10.64
Rate Rate 144.30 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C0
Slope Slope 47.987 ± 0.149 nElectrons/VCal
Offset Offset -135.379 ± 17.71 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C0
Rate Rate 246.30 kHz/cm²
NTrigs N Trig 5913115 Trigger
NHits N Hits 22717 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C0
Rate Rate 387.70 kHz/cm²
NTrigs N Trig 5913243 Trigger
NHits N Hits 35761 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C0
Rate Rate 223.60 kHz/cm²
NTrigs N Trig 5913128 Trigger
NHits N Hits 20622 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C0
Rate Rate 144.30 kHz/cm²
NTrigs N Trig 5912980 Trigger
NHits N Hits 13312 Hits