Test Date: 2015-04-08 11:04
Chip 8
X-ray Calibration - Target Cu Method Spectrum Chip C8
Center Center of Peak 41.02 ± 0.13 Vcal
TargetEnergy Energy of target Cu 8047.82 eV
TargetNElectrons Energy of target Cu 2235.51 nElectrons
Chi2PerNDF Chi^2 per NDF 4.6
Rate Rate 326.20 kHz/cm²
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 43.64 ± 0.16 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 5.26
Rate Rate 343.40 kHz/cm²
X-ray Calibration - Target Br Method Spectrum Chip C8
Center Center of Peak 60.78 ± 0.12 Vcal
TargetEnergy Energy of target Br 11924.36 eV
TargetNElectrons Energy of target Br 3312.32 nElectrons
Chi2PerNDF Chi^2 per NDF 3.83
Rate Rate 542.90 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 93.29 ± 0.12 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.8
Rate Rate 546.30 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 118.93 ± 0.13 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 2.77
Rate Rate 289.30 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 137.23 ± 0.2 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.16
Rate Rate 181.60 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 49.871 ± 0.136 nElectrons/VCal
Offset Offset 215.576 ± 12.965 nElectrons
X-ray Hit Map - Target Cu Method Spectrum Chip C8
Rate Rate 326.20 kHz/cm²
NTrigs N Trig 2956313 Trigger
NHits N Hits 15044 Hits
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 343.40 kHz/cm²
NTrigs N Trig 2956201 Trigger
NHits N Hits 15837 Hits
X-ray Hit Map - Target Br Method Spectrum Chip C8
Rate Rate 542.90 kHz/cm²
NTrigs N Trig 2956067 Trigger
NHits N Hits 25035 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 546.30 kHz/cm²
NTrigs N Trig 2956113 Trigger
NHits N Hits 25195 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 289.30 kHz/cm²
NTrigs N Trig 3750020 Trigger
NHits N Hits 16924 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 181.60 kHz/cm²
NTrigs N Trig 2956395 Trigger
NHits N Hits 8376 Hits