Test Date: 2015-04-23 11:04
Chip 12
X-ray Calibration - Target Cu Method Spectrum Chip C12
Center Center of Peak 47.23 ± 0.09 Vcal
TargetEnergy Energy of target Cu 8047.82 eV
TargetNElectrons Energy of target Cu 2235.51 nElectrons
Chi2PerNDF Chi^2 per NDF 7.36
Rate Rate 307.90 kHz/cm²
X-ray Calibration - Target Zn Method Spectrum Chip C12
Center Center of Peak 49.06 ± 0.11 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 8.13
Rate Rate 337.10 kHz/cm²
X-ray Calibration - Target Br Method Spectrum Chip C12
Center Center of Peak 66.51 ± 0.16 Vcal
TargetEnergy Energy of target Br 11924.36 eV
TargetNElectrons Energy of target Br 3312.32 nElectrons
Chi2PerNDF Chi^2 per NDF 9.14
Rate Rate 540.70 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C12
Center Center of Peak 103.97 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 5.62
Rate Rate 541.40 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C12
Center Center of Peak 132.83 ± 0.14 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 7.21
Rate Rate 290.70 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C12
Center Center of Peak 152.29 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.83
Rate Rate 182.70 kHz/cm²
X-ray Calibration - Target Ba Method Spectrum Chip C12
Center Center of Peak 193.26 ± 0.38 Vcal
TargetEnergy Energy of target Ba 32193.26 eV
TargetNElectrons Energy of target Ba 8942.57 nElectrons
Chi2PerNDF Chi^2 per NDF 3.84
Rate Rate 64.10 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C12
Slope Slope 44.782 ± 0.068 nElectrons/VCal
Offset Offset 201.574 ± 7.457 nElectrons
X-ray Hit Map - Target Cu Method Spectrum Chip C12
Rate Rate 307.90 kHz/cm²
NTrigs N Trig 5911192 Trigger
NHits N Hits 28396 Hits
X-ray Hit Map - Target Zn Method Spectrum Chip C12
Rate Rate 337.10 kHz/cm²
NTrigs N Trig 5911311 Trigger
NHits N Hits 31082 Hits
X-ray Hit Map - Target Br Method Spectrum Chip C12
Rate Rate 540.70 kHz/cm²
NTrigs N Trig 5910896 Trigger
NHits N Hits 49855 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C12
Rate Rate 541.40 kHz/cm²
NTrigs N Trig 5911879 Trigger
NHits N Hits 49935 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C12
Rate Rate 290.70 kHz/cm²
NTrigs N Trig 5912034 Trigger
NHits N Hits 26809 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C12
Rate Rate 182.70 kHz/cm²
NTrigs N Trig 5909390 Trigger
NHits N Hits 16839 Hits
X-ray Hit Map - Target Ba Method Spectrum Chip C12
Rate Rate 64.10 kHz/cm²
NTrigs N Trig 5911628 Trigger
NHits N Hits 5914 Hits