Test Date: 2015-05-26 14:05
Chip 7
X-ray Calibration - Target Zn Method Spectrum Chip C7
Center Center of Peak 38.73 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 19.18
Rate Rate 366.70 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C7
Center Center of Peak 91.39 ± 0.11 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 14.52
Rate Rate 549.90 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C7
Center Center of Peak 116.52 ± 0.14 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 8.7
Rate Rate 290.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C7
Center Center of Peak 135.37 ± 0.23 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 8.06
Rate Rate 181.20 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C7
Slope Slope 50.203 ± 0.253 nElectrons/VCal
Offset Offset 275.11 ± 26.971 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C7
Rate Rate 366.70 kHz/cm²
NTrigs N Trig 5911837 Trigger
NHits N Hits 33817 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C7
Rate Rate 549.90 kHz/cm²
NTrigs N Trig 5911915 Trigger
NHits N Hits 50712 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C7
Rate Rate 290.80 kHz/cm²
NTrigs N Trig 5911947 Trigger
NHits N Hits 26815 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C7
Rate Rate 181.20 kHz/cm²
NTrigs N Trig 5911790 Trigger
NHits N Hits 16708 Hits