Test Date: 2015-03-31 12:03
Chip 8
X-ray Calibration - Target Cu Method Spectrum Chip C8
Center Center of Peak 51.32 ± 0.05 Vcal
TargetEnergy Energy of target Cu 8047.82 eV
TargetNElectrons Energy of target Cu 2235.51 nElectrons
Chi2PerNDF Chi^2 per NDF 1.97
Rate Rate 343.50 kHz/cm²
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 53.18 ± 0.05 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.91
Rate Rate 382.20 kHz/cm²
X-ray Calibration - Target Br Method Spectrum Chip C8
Center Center of Peak 68.73 ± 0.08 Vcal
TargetEnergy Energy of target Br 11924.36 eV
TargetNElectrons Energy of target Br 3312.32 nElectrons
Chi2PerNDF Chi^2 per NDF 2.95
Rate Rate 592.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 99.39 ± 0.06 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.73
Rate Rate 567.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 125.15 ± 0.09 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.62
Rate Rate 302.50 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 142.97 ± 0.12 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.36
Rate Rate 189.10 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 49.925 ± 0.137 nElectrons/VCal
Offset Offset -104.427 ± 15.643 nElectrons
X-ray Hit Map - Target Cu Method Spectrum Chip C8
Rate Rate 343.50 kHz/cm²
NTrigs N Trig 5910172 Trigger
NHits N Hits 31670 Hits
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 382.20 kHz/cm²
NTrigs N Trig 5908882 Trigger
NHits N Hits 35232 Hits
X-ray Hit Map - Target Br Method Spectrum Chip C8
Rate Rate 592.10 kHz/cm²
NTrigs N Trig 3630153 Trigger
NHits N Hits 33529 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 567.10 kHz/cm²
NTrigs N Trig 5910007 Trigger
NHits N Hits 52283 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 302.50 kHz/cm²
NTrigs N Trig 5911383 Trigger
NHits N Hits 27897 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 189.10 kHz/cm²
NTrigs N Trig 5911763 Trigger
NHits N Hits 17443 Hits