Test Date: 2015-03-31 12:03
Chip 2
X-ray Calibration - Target Cu Method Spectrum Chip C2
Center Center of Peak 51.55 ± 0.09 Vcal
TargetEnergy Energy of target Cu 8047.82 eV
TargetNElectrons Energy of target Cu 2235.51 nElectrons
Chi2PerNDF Chi^2 per NDF 1.8
Rate Rate 195.40 kHz/cm²
X-ray Calibration - Target Zn Method Spectrum Chip C2
Center Center of Peak 54.17 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.73
Rate Rate 206.10 kHz/cm²
X-ray Calibration - Target Br Method Spectrum Chip C2
Center Center of Peak 69.61 ± 0.17 Vcal
TargetEnergy Energy of target Br 11924.36 eV
TargetNElectrons Energy of target Br 3312.32 nElectrons
Chi2PerNDF Chi^2 per NDF 3.31
Rate Rate 337.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C2
Center Center of Peak 104.34 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.73
Rate Rate 366.60 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C2
Center Center of Peak 132.18 ± 0.13 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.52
Rate Rate 210.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C2
Center Center of Peak 150.42 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.83
Rate Rate 139.50 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C2
Slope Slope 46.915 ± 0.166 nElectrons/VCal
Offset Offset -40.511 ± 20.229 nElectrons
X-ray Hit Map - Target Cu Method Spectrum Chip C2
Rate Rate 195.40 kHz/cm²
NTrigs N Trig 5910172 Trigger
NHits N Hits 18014 Hits
X-ray Hit Map - Target Zn Method Spectrum Chip C2
Rate Rate 206.10 kHz/cm²
NTrigs N Trig 5908882 Trigger
NHits N Hits 18998 Hits
X-ray Hit Map - Target Br Method Spectrum Chip C2
Rate Rate 337.20 kHz/cm²
NTrigs N Trig 3630153 Trigger
NHits N Hits 19093 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C2
Rate Rate 366.60 kHz/cm²
NTrigs N Trig 5910007 Trigger
NHits N Hits 33796 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C2
Rate Rate 210.80 kHz/cm²
NTrigs N Trig 5911383 Trigger
NHits N Hits 19438 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C2
Rate Rate 139.50 kHz/cm²
NTrigs N Trig 5911763 Trigger
NHits N Hits 12865 Hits