Test Date: 2015-03-31 12:03
Chip 10
X-ray Calibration - Target Cu Method Spectrum Chip C10
Center Center of Peak 52.35 ± 0.15 Vcal
TargetEnergy Energy of target Cu 8047.82 eV
TargetNElectrons Energy of target Cu 2235.51 nElectrons
Chi2PerNDF Chi^2 per NDF 2.13
Rate Rate 394.30 kHz/cm²
X-ray Calibration - Target Zn Method Spectrum Chip C10
Center Center of Peak 54.9 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 2.36
Rate Rate 441.60 kHz/cm²
X-ray Calibration - Target Br Method Spectrum Chip C10
Center Center of Peak 71.2 ± 0.11 Vcal
TargetEnergy Energy of target Br 11924.36 eV
TargetNElectrons Energy of target Br 3312.32 nElectrons
Chi2PerNDF Chi^2 per NDF 4.17
Rate Rate 668.30 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C10
Center Center of Peak 103.15 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.66
Rate Rate 608.80 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C10
Center Center of Peak 130.18 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.7
Rate Rate 314.20 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C10
Center Center of Peak 148.55 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.0
Rate Rate 196.10 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C10
Slope Slope 47.807 ± 0.141 nElectrons/VCal
Offset Offset -74.509 ± 16.794 nElectrons
X-ray Hit Map - Target Cu Method Spectrum Chip C10
Rate Rate 394.30 kHz/cm²
NTrigs N Trig 5910172 Trigger
NHits N Hits 36354 Hits
X-ray Hit Map - Target Zn Method Spectrum Chip C10
Rate Rate 441.60 kHz/cm²
NTrigs N Trig 5908882 Trigger
NHits N Hits 40703 Hits
X-ray Hit Map - Target Br Method Spectrum Chip C10
Rate Rate 668.30 kHz/cm²
NTrigs N Trig 3630153 Trigger
NHits N Hits 37847 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C10
Rate Rate 608.80 kHz/cm²
NTrigs N Trig 5910007 Trigger
NHits N Hits 56131 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C10
Rate Rate 314.20 kHz/cm²
NTrigs N Trig 5911383 Trigger
NHits N Hits 28977 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C10
Rate Rate 196.10 kHz/cm²
NTrigs N Trig 5911763 Trigger
NHits N Hits 18082 Hits