Test Date: 2015-06-04 14:06
Chip 9
X-ray Calibration - Target Zn Method Spectrum Chip C9
Center |
Center of Peak |
52.26 ± 0.12 |
Vcal |
TargetEnergy |
Energy of target Zn |
8638.91 |
eV |
TargetNElectrons |
Energy of target Zn |
2399.7 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
5.34 |
|
Rate |
Rate |
315.90 |
kHz/cm² |
X-ray Calibration - Target Mo Method Spectrum Chip C9
Center |
Center of Peak |
106.44 ± 0.08 |
Vcal |
TargetEnergy |
Energy of target Mo |
17479.37 |
eV |
TargetNElectrons |
Energy of target Mo |
4855.38 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
4.13 |
|
Rate |
Rate |
466.20 |
kHz/cm² |
X-ray Calibration - Target Ag Method Spectrum Chip C9
Center |
Center of Peak |
135.39 ± 0.11 |
Vcal |
TargetEnergy |
Energy of target Ag |
22162.92 |
eV |
TargetNElectrons |
Energy of target Ag |
6156.37 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
4.09 |
|
Rate |
Rate |
245.40 |
kHz/cm² |
X-ray Calibration - Target Sn Method Spectrum Chip C9
Center |
Center of Peak |
153.29 ± 0.13 |
Vcal |
TargetEnergy |
Energy of target Sn |
25271.36 |
eV |
TargetNElectrons |
Energy of target Sn |
7019.82 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
4.42 |
|
Rate |
Rate |
158.10 |
kHz/cm² |
X-ray Calibration - Vcal Calibration Method Spectrum Chip C9
Slope |
Slope |
45.21 ± 0.088 |
nElectrons/VCal |
Offset |
Offset |
39.777 ± 9.291 |
nElectrons |
X-ray Hit Map - Target Zn Method Spectrum Chip C9
Rate |
Rate |
315.90 |
kHz/cm² |
NTrigs |
N Trig |
5913615 |
Trigger |
NHits |
N Hits |
29146 |
Hits |
X-ray Hit Map - Target Mo Method Spectrum Chip C9
Rate |
Rate |
466.20 |
kHz/cm² |
NTrigs |
N Trig |
5913760 |
Trigger |
NHits |
N Hits |
43013 |
Hits |
X-ray Hit Map - Target Ag Method Spectrum Chip C9
Rate |
Rate |
245.40 |
kHz/cm² |
NTrigs |
N Trig |
5913380 |
Trigger |
NHits |
N Hits |
22637 |
Hits |
X-ray Hit Map - Target Sn Method Spectrum Chip C9
Rate |
Rate |
158.10 |
kHz/cm² |
NTrigs |
N Trig |
5913488 |
Trigger |
NHits |
N Hits |
14584 |
Hits |