Test Date: 2015-06-04 14:06
Chip 6
X-ray Calibration - Target Zn Method Spectrum Chip C6
Center Center of Peak 55.88 ± 0.12 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 912.98
Rate Rate 3.30 MHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C6
Center Center of Peak 114.29 ± 0.13 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1016.8
Rate Rate 3.50 MHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C6
Center Center of Peak 145.96 ± 0.14 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 1113.55
Rate Rate 3.30 MHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C6
Center Center of Peak 164.82 ± 0.18 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 1058.24
Rate Rate 3.20 MHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C6
Slope Slope 41.767 ± 0.083 nElectrons/VCal
Offset Offset 69.394 ± 8.894 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C6
Rate Rate 3.30 MHz/cm²
NTrigs N Trig 5913615 Trigger
NHits N Hits 303643 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C6
Rate Rate 3.50 MHz/cm²
NTrigs N Trig 5913760 Trigger
NHits N Hits 320527 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C6
Rate Rate 3.30 MHz/cm²
NTrigs N Trig 5913380 Trigger
NHits N Hits 307109 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C6
Rate Rate 3.20 MHz/cm²
NTrigs N Trig 5913488 Trigger
NHits N Hits 297425 Hits