Test Date: 2015-06-04 14:06
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 49.59 ± 0.22 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 2.71
Rate Rate 257.50 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 101.49 ± 0.11 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.69
Rate Rate 414.40 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 128.68 ± 0.13 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.76
Rate Rate 238.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 147.2 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.82
Rate Rate 152.10 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 47.403 ± 0.116 nElectrons/VCal
Offset Offset 47.677 ± 13.697 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 257.50 kHz/cm²
NTrigs N Trig 5913615 Trigger
NHits N Hits 23754 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 414.40 kHz/cm²
NTrigs N Trig 5913760 Trigger
NHits N Hits 38232 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 238.80 kHz/cm²
NTrigs N Trig 5913380 Trigger
NHits N Hits 22028 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 152.10 kHz/cm²
NTrigs N Trig 5913488 Trigger
NHits N Hits 14027 Hits