Test Date: 2015-03-26 13:03
Chip 11
X-ray Calibration - Target Cu Method Spectrum Chip C11
Center Center of Peak 38.91 ± 0.08 Vcal
TargetEnergy Energy of target Cu 8047.82 eV
TargetNElectrons Energy of target Cu 2235.51 nElectrons
Chi2PerNDF Chi^2 per NDF 4.34
Rate Rate 143.00 kHz/cm²
X-ray Calibration - Target Zn Method Spectrum Chip C11
Center Center of Peak 41.57 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 5.29
Rate Rate 185.00 kHz/cm²
X-ray Calibration - Target Br Method Spectrum Chip C11
Center Center of Peak 58.27 ± 0.07 Vcal
TargetEnergy Energy of target Br 11924.36 eV
TargetNElectrons Energy of target Br 3312.32 nElectrons
Chi2PerNDF Chi^2 per NDF 20.93
Rate Rate 439.50 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C11
Center Center of Peak 86.68 ± 0.06 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 8.66
Rate Rate 520.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C11
Center Center of Peak 111.52 ± 0.08 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 8.69
Rate Rate 283.30 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C11
Center Center of Peak 127.23 ± 0.11 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 7.45
Rate Rate 181.50 kHz/cm²
X-ray Calibration - Target Ba Method Spectrum Chip C11
Center Center of Peak 160.49 ± 0.37 Vcal
TargetEnergy Energy of target Ba 32193.26 eV
TargetNElectrons Energy of target Ba 8942.57 nElectrons
Chi2PerNDF Chi^2 per NDF 2.69
Rate Rate 60.10 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C11
Slope Slope 53.063 ± 0.147 nElectrons/VCal
Offset Offset 252.744 ± 14.967 nElectrons
X-ray Hit Map - Target Cu Method Spectrum Chip C11
Rate Rate 143.00 kHz/cm²
NTrigs N Trig 5911068 Trigger
NHits N Hits 13185 Hits
X-ray Hit Map - Target Zn Method Spectrum Chip C11
Rate Rate 185.00 kHz/cm²
NTrigs N Trig 5909660 Trigger
NHits N Hits 17059 Hits
X-ray Hit Map - Target Br Method Spectrum Chip C11
Rate Rate 439.50 kHz/cm²
NTrigs N Trig 5916923 Trigger
NHits N Hits 40564 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C11
Rate Rate 520.10 kHz/cm²
NTrigs N Trig 5911271 Trigger
NHits N Hits 47962 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C11
Rate Rate 283.30 kHz/cm²
NTrigs N Trig 5911138 Trigger
NHits N Hits 26120 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C11
Rate Rate 181.50 kHz/cm²
NTrigs N Trig 5911046 Trigger
NHits N Hits 16740 Hits
X-ray Hit Map - Target Ba Method Spectrum Chip C11
Rate Rate 60.10 kHz/cm²
NTrigs N Trig 3766905 Trigger
NHits N Hits 3529 Hits