Test Date: 2016-09-19 15:21
Analysis date: 2016-09-19 17:33
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 51.47 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.26
Rate Rate 210.30 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 100.2 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.68
Rate Rate 359.70 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 126.3 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.82
Rate Rate 209.40 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 143.75 ± 0.12 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 5.23
Rate Rate 136.70 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 49.737 ± 0.147 nElectrons/VCal
Offset Offset -127.789 ± 17.126 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 210.30 kHz/cm²
NTrigs N Trig 5914131 Trigger
NHits N Hits 19403 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 359.70 kHz/cm²
NTrigs N Trig 5914224 Trigger
NHits N Hits 33191 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 209.40 kHz/cm²
NTrigs N Trig 5914334 Trigger
NHits N Hits 19318 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 136.70 kHz/cm²
NTrigs N Trig 5914010 Trigger
NHits N Hits 12610 Hits
MoReWeb-v1.0.2-10-gbc52a17 on branch psi46master