Test Date: 2016-08-10 16:30
Analysis date: 2016-08-12 11:19
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 52.12 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.28
Rate Rate 216.90 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 104.88 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.72
Rate Rate 375.30 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 132.46 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.9
Rate Rate 218.30 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 150.23 ± 0.12 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 5.19
Rate Rate 144.10 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 46.714 ± 0.075 nElectrons/VCal
Offset Offset -37.787 ± 7.56 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 216.90 kHz/cm²
NTrigs N Trig 5916287 Trigger
NHits N Hits 20020 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 375.30 kHz/cm²
NTrigs N Trig 5916292 Trigger
NHits N Hits 34638 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 218.30 kHz/cm²
NTrigs N Trig 5916483 Trigger
NHits N Hits 20152 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 144.10 kHz/cm²
NTrigs N Trig 5916650 Trigger
NHits N Hits 13300 Hits
MoReWeb-v1.0.2-10-gbc52a17 on branch psi46master