Test Date: 2016-06-29 10:04
Analysis date: 2016-06-29 11:49
Chip 10
X-ray Calibration - Target Zn Method Spectrum Chip C10
Center Center of Peak 50.09 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.32
Rate Rate 315.40 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C10
Center Center of Peak 98.36 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.53
Rate Rate 460.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C10
Center Center of Peak 123.96 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 2.68
Rate Rate 235.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C10
Center Center of Peak 140.93 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 2.99
Rate Rate 147.30 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C10
Slope Slope 50.838 ± 0.176 nElectrons/VCal
Offset Offset -145.081 ± 20.16 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C10
Rate Rate 315.40 kHz/cm²
NTrigs N Trig 5916309 Trigger
NHits N Hits 29106 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C10
Rate Rate 460.10 kHz/cm²
NTrigs N Trig 5916325 Trigger
NHits N Hits 42467 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C10
Rate Rate 235.80 kHz/cm²
NTrigs N Trig 5916490 Trigger
NHits N Hits 21768 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C10
Rate Rate 147.30 kHz/cm²
NTrigs N Trig 5916284 Trigger
NHits N Hits 13595 Hits
MoReWeb-v1.0.1-7-g5388ca4 on branch psi46master