Test Date: 2016-06-15 15:42
Analysis date: 2016-06-15 17:27
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 47.47 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.18
Rate Rate 213.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 99.37 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.47
Rate Rate 367.70 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 126.14 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 2.95
Rate Rate 211.10 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 143.87 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.63
Rate Rate 137.50 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 48.628 ± 0.175 nElectrons/VCal
Offset Offset 23.102 ± 20.552 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 213.10 kHz/cm²
NTrigs N Trig 5916213 Trigger
NHits N Hits 19668 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 367.70 kHz/cm²
NTrigs N Trig 5916222 Trigger
NHits N Hits 33938 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 211.10 kHz/cm²
NTrigs N Trig 5916153 Trigger
NHits N Hits 19487 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 137.50 kHz/cm²
NTrigs N Trig 5916291 Trigger
NHits N Hits 12694 Hits