Test Date: 2016-06-10 10:53
Analysis date: 2016-06-10 13:57
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 52.88 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.38
Rate Rate 225.80 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 104.52 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.61
Rate Rate 384.90 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 131.72 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.81
Rate Rate 218.60 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 150.12 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.99
Rate Rate 143.50 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 47.625 ± 0.086 nElectrons/VCal
Offset Offset -120.205 ± 8.695 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 225.80 kHz/cm²
NTrigs N Trig 5913957 Trigger
NHits N Hits 20828 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 384.90 kHz/cm²
NTrigs N Trig 5913705 Trigger
NHits N Hits 35509 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 218.60 kHz/cm²
NTrigs N Trig 5913973 Trigger
NHits N Hits 20166 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 143.50 kHz/cm²
NTrigs N Trig 5913786 Trigger
NHits N Hits 13237 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master