Test Date: 2016-06-10 10:53
Analysis date: 2016-06-10 13:53
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 51.81 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.56
Rate Rate 331.80 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 102.54 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.61
Rate Rate 479.20 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 129.31 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 2.7
Rate Rate 252.30 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 146.69 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.0
Rate Rate 160.30 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 48.45 ± 0.078 nElectrons/VCal
Offset Offset -111.201 ± 7.356 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 331.80 kHz/cm²
NTrigs N Trig 5916532 Trigger
NHits N Hits 30624 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 479.20 kHz/cm²
NTrigs N Trig 5916406 Trigger
NHits N Hits 44230 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 252.30 kHz/cm²
NTrigs N Trig 5916536 Trigger
NHits N Hits 23286 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 160.30 kHz/cm²
NTrigs N Trig 5916775 Trigger
NHits N Hits 14793 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master