Test Date: 2016-05-12 13:37
Analysis date: 2016-05-26 17:15
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 53.73 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.35
Rate Rate 267.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 107.3 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.2
Rate Rate 407.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 135.54 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.52
Rate Rate 232.60 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 154.24 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 5.91
Rate Rate 150.90 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 46.099 ± 0.157 nElectrons/VCal
Offset Offset -91.192 ± 19.868 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 267.10 kHz/cm²
NTrigs N Trig 5917187 Trigger
NHits N Hits 24660 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 407.50 kHz/cm²
NTrigs N Trig 5916689 Trigger
NHits N Hits 37608 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 232.60 kHz/cm²
NTrigs N Trig 5917188 Trigger
NHits N Hits 21474 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 150.90 kHz/cm²
NTrigs N Trig 5917106 Trigger
NHits N Hits 13928 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master