Test Date: 2016-05-04 10:00
Analysis date: 2016-05-26 16:38
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 52.31 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.45
Rate Rate 311.90 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 101.73 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.38
Rate Rate 455.40 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 128.9 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.45
Rate Rate 236.20 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 146.8 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.45
Rate Rate 152.80 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 47.983 ± 0.142 nElectrons/VCal
Offset Offset -26.353 ± 16.736 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 311.90 kHz/cm²
NTrigs N Trig 5913853 Trigger
NHits N Hits 28773 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 455.40 kHz/cm²
NTrigs N Trig 5913830 Trigger
NHits N Hits 42016 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 236.20 kHz/cm²
NTrigs N Trig 5913690 Trigger
NHits N Hits 21791 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 152.80 kHz/cm²
NTrigs N Trig 5913757 Trigger
NHits N Hits 14093 Hits