Test Date: 2016-05-04 10:00
Analysis date: 2016-05-26 16:38
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 49.2 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.35
Rate Rate 222.60 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 102.12 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.71
Rate Rate 373.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 129.43 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.81
Rate Rate 215.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 147.37 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 5.65
Rate Rate 138.70 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 47.777 ± 0.169 nElectrons/VCal
Offset Offset -24.374 ± 20.358 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 222.60 kHz/cm²
NTrigs N Trig 5913853 Trigger
NHits N Hits 20535 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 373.50 kHz/cm²
NTrigs N Trig 5913830 Trigger
NHits N Hits 34454 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 215.80 kHz/cm²
NTrigs N Trig 5913690 Trigger
NHits N Hits 19911 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 138.70 kHz/cm²
NTrigs N Trig 5913757 Trigger
NHits N Hits 12794 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master