Test Date: 2016-04-20 09:33
Analysis date: 2016-05-26 15:42
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 51.32 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.42
Rate Rate 203.40 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 102.48 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.06
Rate Rate 363.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 129.78 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.29
Rate Rate 210.20 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 147.71 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.56
Rate Rate 136.40 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 47.799 ± 0.156 nElectrons/VCal
Offset Offset -43.781 ± 18.63 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 203.40 kHz/cm²
NTrigs N Trig 5913757 Trigger
NHits N Hits 18764 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 363.50 kHz/cm²
NTrigs N Trig 5914046 Trigger
NHits N Hits 33532 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 210.20 kHz/cm²
NTrigs N Trig 5913793 Trigger
NHits N Hits 19392 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 136.40 kHz/cm²
NTrigs N Trig 5913770 Trigger
NHits N Hits 12585 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master