Test Date: 2016-02-26 10:52
Analysis date: 2016-05-26 11:45
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 55.36 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 19.31
Rate Rate 342.80 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 110.25 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 28.86
Rate Rate 508.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 139.06 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 23.88
Rate Rate 283.50 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 157.95 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 22.18
Rate Rate 193.70 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 45.309 ± 0.138 nElectrons/VCal
Offset Offset -140.814 ± 17.839 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 342.80 kHz/cm²
NTrigs N Trig 5916401 Trigger
NHits N Hits 31642 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 508.50 kHz/cm²
NTrigs N Trig 5916451 Trigger
NHits N Hits 46930 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 283.50 kHz/cm²
NTrigs N Trig 5916302 Trigger
NHits N Hits 26165 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 193.70 kHz/cm²
NTrigs N Trig 5916300 Trigger
NHits N Hits 17877 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master